Technical data
220 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
12 Marker Fctn
Select Marker
Selects one of the four possible marker or marker pairs. Once a marker is selected, it can be
set to any of the control modes, Normal, Delta, Delta Pair, Span Pair, or Off.
Key Path: Marker
State Saved: The number of the selected marker is saved in Instrument State.
Factory Preset: Marker 1
Remote Command:
:CALCulate:MARKer[1]|2|3|4:STATe OFF|ON|0|1
:CALCulate:MARKer[1]|2|3|4:STATe?
Sets or queries the state of a marker. Setting a marker to state ON or 1 selects that marker.
Setting a marker which is OFF to state ON or 1 puts it in Normal mode and places it at the
center of the display. Setting a marker to state OFF or 0 selects that marker and turns it off.
The response to the query will be 0 if OFF, 1 if ON.
Example:
CALC:MARK2:STAT ON selects marker 2.
CALC:MARK:STAT ON will not modify a marker that is already on.