Technical data

214 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
11 Marker
Period
Sets the marker readout to Period, displaying the reciprocal of the frequency at the marker
position, or the reciprocal of the frequency separation of the two markers in a delta-marker
mode. Period readout is not available in zero spans. If the markers are at the same
frequency in a delta marker mode, the result will be the reciprocal of 0, which is infinitely
large. The display will show a very large number.
Key Path: Marker, Readout
Remote Command:
See“Readout" on page 213 for this command.
Example:
CALC:MARK2:X:READ PER
Time
Sets the marker readout to Time, displaying the time interval between a normal marker and
the start of a sweep or the time of the delta marker relative to the reference marker. Time is
the default setting in zero spans. With a span of zero, the time value is the time position
relative to the start of the sweep. In a delta-marker mode it is the (sweep) time interval
between the two markers.
Key Path: Marker, Readout
Remote Command:
See“Readout" on page 213 for this command.
Example:
CALC:MARK2:X:READ TIME
Inverse Time
Sets the marker readout to Inverse Time, displaying the reciprocal of (sweep) time between
two markers. This function is only available when in both zero span and in a delta-marker
modes. If the markers are at the same x position, the time between them is 0, so the
reciprocal of sweep time is infinitely large. The display will show a very large number.
Key Path: Marker, Readout
Remote Command:
See “Readout" on page 213 for this command.