Technical data

Marker 11
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 213
Readout
Access a menu that enables you to affect how the x-axis information for the selected
marker is displayed in the marker area (top-right of display) and the active function area of
the display. It only affects the readout on the display of the horizontal position information
(for example, frequency).
It does not affect the way this information is sent remotely in response to the
CALC:MARK:X? command.
Key Path: Marker
State Saved: In Instrument State, for each marker.
Factory Preset: Frequency for non-zero spans and time for zero spans.
Remote Command:
:CALCulate:MARKer[1]|2|3|4:X:READout
FREQuency|TIME|ITIMe|PERiod
:CALCulate:MARKer[1]|2|3|4:X:READout?
Example:
CALC:MARK3:X:READ TIME sets the marker 3 Readout to Time.
Frequency
Sets the marker readout to Frequency, displaying the absolute frequency of a normal marker
or the frequency of the delta marker relative to the reference marker. Frequency readout is
the default setting in non-zero spans and is not available in zero spans.
Key Path: Marker, Readout
Remote Command:
See“Readout" on page 213 for this command.
Example:
CALC:MARK2:X:READ FREQ sets the marker 2 Readout to Frequency.