Technical data
Input/Output 9
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 193
Image Shift
Performs signal identification in a two sweep sequence. Places data from the first sweep in
Trace 1, and data from the second (frequency shifted) sweep in Trace 2. Signal responses of
Trace 1 and Trace 2 having the same horizontal position are considered to be in the current
band and therefore can be analyzed with the amplitude and frequency measurement
systems of the N8201A. All other responses are invalid and should be ignored.
Key Path: Input/Output, External Mixer, Signal ID Mode
State Saved: Saved in Instrument State
Remote Command: See “Signal ID Mode" on page 192.
Example: SID:MODE ISH
Mixer Config
Displays the Mixer Config menu keys to manually set the harmonic, select the preselected
mixers or unpreselected mixers, and adjust the internal bias source for use with mixers
requiring bias.
Key Path: Input/Output, External Mixer
State Saved: Saved in Instrument State
Harmonic
The harmonic mixing number, its associated sign, and the availability of mixer bias can be
automatically controlled by setting Harmonic to Auto. In Auto, the harmonic number and
sign are determined by the Ext Mix Band selected. For bands A, Q, U and V, they are
determined by the Ext Mix Band and by whether Mixer Type is set to preselected or
unpreselected. There are no auto rules for Ext Mix Band set to Auto, therefore, selecting
Auto forces Ext Mix Band to A band.
The harmonic number indicated is a signed number. Positive numbers (sign not displayed)
indicate that the tuned frequency is above the desired LO harmonic by the 321.4 MHz IF.
Negative numbers indicate an LO harmonic below the tuned frequency by the 321.4 MHz IF.
Key Path: Input/Output, External Mixer, Mixer Config
State Saved: Saved in Instrument State
Factory Preset: Auto
Range: –50 to 50, cannot be set to 0.
Remote Command: