Technical data

File 7
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 157
Complete path
A file extension of TRC
Use all uppercase letters.
Trace labels are: TRACE1|TRACE2|TRACE3
For the MMEM:LOAD:STAT 1 command:
If the firmware revision of the state being loaded is newer than the firmware revision of
the instrument, no state is recalled and an error is reported.
If the firmware revision of the state being loaded is the same as the firmware revision of
the instrument, all settings of the state will be loaded.
If the firmware revision of the state being loaded is older than the firmware revision of
the instrument, the instrument will only load the older settings of the state.
Example:
MMEM:LOAD:STAT 1,’C:MYSTATE.STA’ loads the state file C:\
MYSTATE.STA.
MMEM:LOAD:TRAC TRACE3,’C:MYTRACE.TRC’ loads the trace in file C:\
MYTRACE.TRC into trace 3.
Type
Enables you to select the type of file you want to load. See “File Types" on page 144 and
“Type" on page 146 for more information.
The file types available for loading are described in the following table:
All is not a file type option in Load. You have to specify the desired file type.
Key Path: File, Load
Sort
Allows you to view saved files according to a selected file attribute. See “Sort" on page 148
for more information.
Type Format Destination Extension
State State STA
Trac e Trace + state Trace 1, 2, 3, or all traces TRC
Comma separated trace values Trace 1, 2, 3, or all traces CSV
Limit Internal data format LIM
Corrections Internal data format ANT, CBL, OTH, and AMP