Technical data

152 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
7 File
The file name must have a file extension of .TRC or .CSV. The file extension determines
whether a trace is stored (.CSV), or a trace with its state (.TRC), are stored.
The <‘file_name’> must include the complete path, for example ‘C:\
MYTRACE.TRC’. Lowercase characters are interpreted as uppercase.
These commands will fail if the
<‘file_name’> already exists.
For the MMEM:STOR:LIM LLINE2,”C:mylimit.lim” command, there is no short
form for parameters LLINE1|LLINE2.
Example:
MMEM:STOR:STAT 1,’C:\mystate.sta’ saves the current instrument state to
the specified file name. The .sta extension is required.
*SAV saves the current instrument state to a file name REGxxx, where xxx = the
register number. The available register numbers are 0 to 127.
MMEM:STOR:SCR ‘C:\myscreen.gif’ The file must have a .gif file extension.
The specified file extension determines which file format the instrument will use to save
the image. Only
Bitmap is available (not Reverse Bitmap).
MMEM:STOR:TRAC TRACE3,’C:\mytrace.trc’ Saves trace 3 to the trace +
state file C:\MYTRACE.TRC
MMEM:STOR:CORR ANT, ‘A:TEST1.AMP’ saves the current antenna correction to the
specified file name. The .amp extension is required.
MMEM:STOR:LIM LLINE2,’C:mylimit.lim’ saves the current limit line two data
set to the specified file name. The .lim extension is required.
MMEM:STOR:RES ‘A:ACP.CSV’ saves the current ACP measurement results to the
specified file name. The .csv extension is required.
Type
Allows you to select the type of data you want to save. The file types available for saving
are described below. See “File Types" on page 144 and “Type" on page 146 for more
information.
Type Format Source Extension
State State STA
Trac e Trace + state Trace 1, 2, 3, or all traces TRC
Comma separated trace values Trace 1, 2, 3, or all traces CSV
Limit Internal data format LIM
Screen Bitmap GIF
Reverse bitmap GIF