Technical data
File 7
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 151
Never remove the floppy disk during a save operation. To do so could corrupt all data on the
floppy disk.
Many errors can be generated by a bad
Save operation. For this reason, if an ‘Unable to
Save file’ message is seen, you should check the error queue (
System, Show Errors)
for the source of the error.
You are always safe pressing
Save Now without entering a file name, because the
auto-generated file name never conflicts with an existing file.
If the Path: field above the directory box is empty when pressing
Save Now, the status
line will display the error message: Unable to save file, invalid path. In this
case, please select a drive.
Key Path: File, Save
Remote Command:
:MMEMory:STORe:SCReen <‘file_name’>
:MMEMory:STORe:STATe 1,<‘file_name’>
The MMEM:STOR:STAT command only saves the state of the Signal Analysis mode. Use
*SAV to save all mode states.
:MMEMory:STORe:TRACe <label>,<‘file_name’>
This command is not available when you are outside the Spectrum Analysis mode and
working in optional measurement modes.
:MMEMory:STORe:LIMit LLINE1|LLINE2,<‘file_name’>
:MMEMory:STORe:CORRection
ANTenna|CABLe|OTHer|USER,<‘file_name’>
:MMEMory:STORe:RESults <‘file_name’>
*SAV <register#>
Remote Command Notes:
For the MMEM:STOR:TRAC <label>,<“file_name”> command:
Trace labels are: TRACE1|TRACE2|TRACE3|ALL
Type Auto-Generated
File Name
Extension
State STATE### .STA
Trac e TRACE### .TRC or
.CSV
Screen SCREN### .GIF