Technical data

Display 6
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 141
Title
Displays menu keys that enable you to change or clear a title on your display.
Key Path: Display
Change Title
Allows you to write a title across the top of the display. Press Change Title to access the
Alpha Editor Menus that contain available characters and symbols. You may also use the
numeric keypad to enter numbers. Press
Enter or Return to complete the entry. Press ESC
to cancel the entry and preserve your existing title.
The display title will remain until you press
Change Title again, or you recall a trace or
state, or a
Factory Preset is performed. A title can also be cleared by pressing Title, Clear
Title
.
Pressing this key cancels any active function.
Key Path: Display, Title
State Saved: Saved in Instrument State.
Remote Command:
:DISPlay:ANNotation:TITLe:DATA <string>
:DISPlay:ANNotation:TITLe:DATA?
Example:
DISP:ANN:TITL:DATA “This Is My Title”
Clear Title
Allows you to clear a title from the front-panel display. Once cleared, the title cannot be
retrieved.
Key Path: Display, Title
Factory Preset: No title
Remote Command:
There is no equivalent command, but the example below shows how to enter an empty title.
Example:
DISP:ANN:TITL:DATA “TEST 3”