Technical data
128 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
6 Display
Limit Display
Turns limit-line display On or Off. Either Limit or Margin, as well as Test, must be turned on
to turn on a limit test.
Key Path:
Display, Limits, Limit 1
Display, Limits, Limit 2
State Saved: Not saved in instrument state.
Factory Default: Off (when a limit line is created)
Remote Command:
:CALCulate:LLINe[1]|2:DISPlay OFF|ON|0|1
:CALCulate:LLINe[1]|2:DISPlay?
Example:
:CALC:LLIN2:DISP OFF
:CALC:LLIN1:DISP?
Limit Test
Turns the testing of the limit line On or Off. If the trace is at or within the bounds of the set
limit or margin, PASS LIMIT# or PASS MARGIN# is displayed in green in the upper-left
corner of the measurement area where # is the number of the selected limit line.
Only positive margins are allowed for lower limits and only negative margins are allowed for
upper limits. If the trace is out of the limit or margin boundaries, FAIL LIMIT# or FAIL
MARGIN# is displayed in red. The results for Limit 2 are displayed below those for Limit 1.
Either Limit or Margin must be turned on for Test to be turned on.
Key Path:
Display, Limits, Limit 1
Display, Limits, Limit 2
Dependencies/Couplings: If either of the limits or margins are turned off, the test
cannot be turned on. That is, if both Limit and Margin are set to Off, then the test is turned
off automatically.
State Saved: Not saved in instrument state.
Factory Default: Off (when a limit line is created)