Technical data
Display 6
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 127
Remote Command Notes:
Up to 200 points total may be defined for each limit.
Example:
CALC:LLIN1:DATA 1000000000, -20,0,200000000,-30,1
Limit 1 or Limit 2
Selects Limit 1 or Limit 2 for modification.
Key Path: Display, Limits
State Saved: Not saved in instrument state.
Type (Upper Lower)
Defines the limit you are editing as either an upper or lower limit. An Upper limit fails if the
trace exceeds the limit. A Lower limit fails if the trace falls below the limit.
Key Path:
Display, Limits, Limit 1
Display, Limits, Limit 2
Dependencies/Couplings: If a margin has already been set for this limit line, and this
key is used to change the limit type, then the margin value is reset to 0 dB.
State Saved: Not saved in instrument state. Survives power cycle and preset. The
limit-line data is saved in files.
Factory Default: Limits are off by default. Upper for Limit 1, then a limit line is created.
Lower for Limit 2, then a limit line is created.
Remote Command:
:CALCulate:LLINe[1]|2:TYPE UPPer|LOWer
:CALCulate:LLINe[1]|2:TYPE?
Example:
:CALC:LLIN2:TYPE LOW sets limit line 2 as a lower limit.
:CALC:LLIN1:TYPE? responds with the limit line 1 limit
type.