Technical data
120 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
5 Det/Demod
MIL Peak
Selects peak detection. This is the same detection as the regular Peak detector. The
difference is the resolution bandwidths that are available, as shown in Table 4 on page 120
and Tab le 5 on page 120.
Key Path: Det/Demod, Detector
Table 4 MIL Specifications for Bandwidth vs. Frequency with RBW Set to Auto
Start Frequency Stop Frequency Bandwidth
30 Hz 1 kHz 10 Hz
1 kHz 10 kHz 100 Hz
10 kHz 150 kHz 1 kHz
150 kHz 30 MHz 10 kHz
30 MHz 1 GHz 100 kHz
1 GHz Unlimited 1 MHz
Table 5 MIL Bandwidths with RBW Set to Manual
RBW, Hz Type Equivalent
− 3 dB BW Filter
10 MIL 6.8 Hz
30 − 6 dB 20 Hz
100 MIL 68 Hz
300 − 6 dB 200 Hz
1 k MIL 680 Hz
3 k − 6 dB 2 kHz
10 k MIL 6.8 kHz
30 k − 6 dB 20 kHz
100 k MIL 68 kHz
300 k − 6 dB 200 kHz
1 M MIL 680 kHz
3 M − 6 dB 2 MHz
10 M − 6 dB 8 MHz