Technical data

314 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
10 Digital Modulation Commands
Digital Modulation Subsystem ([:SOURce]:DM)
:EXTernal:Filter
Supported N8212A
[:SOURce]:DM:EXTernal:FILTer 40e6|THRough
[:SOURce]:DM:EXTernal:FILTer?
Selects the filter or through path for I/Q signals routed to the I and Q outputs.
40e6 Applies a 40 MHz baseband filter.
THRough Bypasses filtering.
Example
:DM:EXT:FILT 40E6
The example selects the 40 MHz baseband filter.
*RST THR
:EXTernal:Filter:AUTO
Supported N8212A
[:SOURce]:DM:EXTernal:FILTer:AUTO ON|OFF|1|0
[:SOURce]:DM:EXTernal:FILTer:AUTO?
Enables or disables the automatic filter selection for I/Q signals routed to the rear-panel
I/Q outputs.
ON(1) Automatically selects the 40 MHz filter optimized for the current N8211A/N8212A
settings.
OFF(0) Disables the auto feature and allows you to select the 40 MHz filter or a through
path.
Example
:DM:EXT:FILT:AUTO 1
The example allows automatic selection of the 40 MHz I/Q filter.
*RST 1 (ON)