Technical data
Analog Commands 9
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 303
IMMediate Enables immediate triggering of the sweep event.
EXTernal Enables the triggering of a sweep event by an externally applied signal at the
TRIGGER IN connector.
Refer to “:PM[1]|2:INTernal[1]:FUNCtion:SHAPe" on page 301 for the waveform selection.
Example
:PM2:INT:SWE:TRIG BUS
The example selects a BUS trigger as the triggering for the internal source 1 swept–sine
waveform on ΦM path 2.
*RST IMM
:PM[1]|2:SOURce
Supported Option UNT
[:SOURce]:PM[1]|2:SOURce INT[1]|INT2|EXT[1]|EXT2
[:SOURce]:PM[1]|2:SOURce?
Selects the source used to generate the phase modulation.
INT Selects internal source 1 or internal source 2 to provide an ac-coupled signal.
EXT Selects the EXT 1 INPUT or the EXT 2 INPUT connector to provide an externally
applied signal that can be ac- or dc- coupled.
Example
:PM2:SOUR EXT1
The example selects an external signal on the EXT 1 INPUT connector as the source for ΦM
path 2 modulation.
*RST INT
:PM[1]|2:STATe
Supported Option UNT
[:SOURce]:PM[1]|2:STATe ON|OFF|1|0
[:SOURce]:PM[1]|2:STATe?