The example selects a positive ramp slope for modulating the signal on ΦM path 1, internal
source 2.
*RST POS
:PM[1]|2:INTernal[1]:FREQuency:ALTernate:AMPLitude:PERCent
Supported Option UNT
[:SOURce]:PM[1]|2:INTernal[1]:FREQuency:ALTernate:AMPLitude:PERCent <val>
[:SOURce]:PM[1]|2:INTernal[1]:FREQuency:ALTernate:AMPLitude:PERCent?
Sets the amplitude of the second tone for the dual–sine waveform as a percentage of the
total amplitude. For example, if the second tone makes up 30% of the total amplitude, then
the first tone is 70% of the total amplitude. Refer to
“:PM[1]|2:INTernal[1]:FUNCtion:SHAPe" on page 301 for the waveform selection.
Example
:PM2:INT:FREQ:ALT:AMPL:PERC 40
The example sets the alternate tone amplitude to 40% of the total amplitude.
*RST +5.00000000E+001
Range 0–100%
:PM[1]|2:INTernal[1]:FUNCtion:SHAPe
Supported Option UNT
[:SOURce]:PM[1]|2:INTernal[1]:FUNCtion:SHAPe
SINE|TRIangle|SQUare|RAMP|NOISe|DUALsine|SWEPtsine
[:SOURce]:PM[1]|2:INTernal[1]:FUNCtion:SHAPe?
Sets the phase modulation waveform type for internal source 1.
Example
:PM1:INT:FUNC:SHAP RAMP
The example selects a ramp modulation for ΦM path 1, source 1.
*RST SINE