Technical data
298 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
9 Analog Commands
The example sets the step value to 100 kHz.
Range 0.5–1E6
:PM[1]|2:BANDwidth|BWIDth
Supported Option UNT
[:SOURce]:PM[1]|2:BANDwidth|BWIDth NORMal|HIGH
[:SOURce]:PM[1]|2:BANDwidth|BWIDth?
Selects normal phase modulation or high bandwidth phase modulation. The command can
use either the BANDwidth or BWIDth paths.
Example
:PM1:BAND NORM
The example selects normal phase modulation for ΦM path 1.
*RST NORM
:PM[1]|2:EXTernal[1]|2:COUPling
Supported Option UNT
[:SOURce]:PM[1]|2:EXTernal[1]|2:COUPling AC|DC
[:SOURce]:PM[1]|2:EXTernal[1]|2:COUPling?
Sets the coupling for the phase modulation source at the selected external input connector.
AC Passes ac signal components.
DC Passes both ac and dc signal components.
This command does not change the active source or switch modulation on or off. The
modulating signal may be the sum of several signals, from either internal or external
sources.
Example
:PM1:EXT:COUP AC
The example selects AC coupling at the external input for ΦM path 1.
*RST DC