Technical data
294 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
9 Analog Commands
Selects a positive or negative slope for the ramp modulation on the LF output.
Refer to “:LFOutput:FUNCtion[1]|2:SHAPe" on page 293 for selecting the waveform type.
Example
:LFO:FUNC1:SHAP:RAMP POS
The example selects a positive ramp slope modulation for the Function Generator 1 LF
output.
*RST POS
:LFOutput:FUNCtion[1]:SWEep:RATE
Supported Option UNT
[:SOURce]:LFOutput:FUNCtion[1]:SWEep:RATE <val><units>
[:SOURce]:LFOutput:FUNCtion[1]:SWEep:RATE?
Sets the sweep rate for an internally generated swept–sine signal.
Example
:LFO:FUNC1:SWE:RATE 1E5
The example sets the sweep rate for the swept–sine waveform to 100 kHz.
*RST +4.00000000E+002
Range 0.5 Hz–100 kHz
:FUNCtion[1]:SWEep:TRIGger
Supported Option UNT
[:SOURce]:LFOutput:FUNCtion[1]:SWEep:TRIGger BUS|IMMediate|EXTernal
[:SOURce]:LFOutput:FUNCtion[1]:SWEep:TRIGger?
Sets the trigger source for the internally generated swept–sine signal at the LF output.
BUS Enables LAN triggering using the *TRG command.
IMMediate Enables immediate triggering of the sweep event.
EXTernal Enables the triggering of a sweep event by an externally applied signal at the
TRIGGER IN connector.
Refer to “:LFOutput:FUNCtion[1]|2:SHAPe" on page 293 for selecting the waveform type.