Technical data
292 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
9 Analog Commands
:LFOutput:FUNCtion[1]:FREQuency:ALTernate
Supported Option UNT
[:SOURce]:LFOutput:FUNCtion[1]:FREQuency:ALTernate <val><units>
[:SOURce]:LFOutput:FUNCtion[1]:FREQuency:ALTernate?
Sets the frequency for the alternate LF output signal. The alternate frequency is the second
tone of a dual–sine or the stop frequency of a swept–sine waveform.
Refer to “:LFOutput:FUNCtion[1]|2:SHAPe" on page 293 for more information on selecting
the waveform type.
Example
:LFO:FUNC1:FREQ:ALT 20KHZ
The example sets the alternate frequency to 20 kHz.
*RST +4.00000000E+002
Range Dual Sine: 0.1 Hz - 100 kHz Swept-Sine: 0.1 Hz - 100 kHz
:LFOutput:FUNCtion[1]:FREQuency:ALTernate:AMPLitude:PERCent
Supported Option UNT
[:SOURce]:LFOutput:FUNCtion[1]:FREQuency:ALTernate:AMPLitude:PERCent
<val><units>
[:SOURce]:LFOutput:FUNCtion[1]:FREQuency:ALTernate:AMPLitude:PERCent?
Sets the amplitude of the second tone for a dual–sine waveform as a percentage of the total
LF output amplitude. For example, if the second tone makes up 30% of the total amplitude,
then the first tone is 70% of the total amplitude. Refer to
“:LFOutput:FUNCtion[1]|2:SHAPe" on page 293 for selecting the waveform type.
Example
:LFO:FUNC1:FREQ:ALT:AMPL:PERC 50
The example sets the alternate frequency to 50% of the total output amplitude.
*RST +5.00000000E+001
Range 0–100%