Technical data
290 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
9 Analog Commands
*RST +1.00000000E+003
:FM[1]|2[:DEViation]:TRACk
Supported Option UNT
[:SOURce]:FM[1]|2[:DEViation]:TRACk ON|OFF|1|0
[:SOURce]:FM[1]|2[:DEViation]:TRACk?
Enables or disables deviation coupling between FM paths 1 and 2.
ON (1) Links the deviation value of FM1 with FM2; FM2 will assume the FM1 deviation
value. For example, if FM1 deviation is set to 500 Hz and FM2 is set to 2 kHz, enabling the
deviation tracking will cause the FM2 deviation value to change to 500 Hz. This applies
regardless of the path (FM1 or FM2) selected.
OFF (0) Disables the coupling and both paths will have independent deviation values.
This command uses exact match tracking, not offset tracking.
Example
:FM2:TRAC 0
The example disables deviation coupling.
*RST 0
Range Frequency Deviation
250 kHz−250ΜHZ 0–2 MHz
> 250−500 MHz 0–1 MHz
> 0.5−1 GHz 0–2 MHz
> 1−2 GHz 0–4 MHz
> 2−3.2 GHz 0–8 MHz
> 3.2−10 GHz 0–16 MHz
> 10−20 GHz 0–32 MHz
> 20−28.5 GHz
*
* N8211A Option 540 only
0–48MHZ
> 20−40 GHz
∗
0–64 MHz