Technical data

Analog Commands 9
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 289
Example
:FM2:SOUR INT2
The example selects internal source 2 as the FM source for FM path 2.
*RST INT
:FM[1]|2:STATe
Supported Option UNT
[:SOURce]:FM[1]|2:STATe ON|OFF|1|0
[:SOURce]:FM[1]|2:STATe?
Enables or disables the selected FM path.
The RF carrier is modulated when you set the modulation state to ON, see
“:MODulation[:STATe]" on page 196 for more information.
The two paths for frequency modulation can be simultaneously enabled. Refer to
“:FM[1]|2" on page 283 for more information.
Example
:FM2:STAT ON
The example enables FM path 2.
*RST 0
:FM[1]|2[:DEViation]
Supported Option UNT
[:SOURce]:FM[1]|2[:DEViation] <val><units>
[:SOURce]:FM[1]|2[:DEViation]?
Sets the FM deviation for the selected FM path.
If deviation tracking is ON, a change to the deviation value on one path will apply to both.
Refer to “:FM[1]|2[:DEViation]:TRACk" on page 290 for more information on setting the
deviation tracking.
Example
:FM2 1MHZ
The example sets the frequency deviation to 1 MHz.