Technical data

278 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
9 Analog Commands
[:SOURce]:AM[1]|2:INTernal[1]:SWEep:RATE?
Sets the sweep rate for the AM swept–sine waveform.
Refer to:AM[1]|2:INTernal[1]|2:FUNCtion:SHAPe" on page 277 for the waveform
selection. The sweep rate function is only available for internal source 1.
Example
:AM2:INT1:SWE:RATE 1KHZ
The example sets the sweep rate for AM path 1, internal source 1 to 1 kHz.
*RST +4.00000000E+002
Range 0.5 Hz–100 kHz
:AM[1]|2:INTernal[1]:SWEep:TRIGger
Supported Option UNT
[:SOURce]:AM[1]|2:INTernal[1]:SWEep:TRIGger BUS|IMMediate|EXTernal
[:SOURce]:AM[1]|2:INTernal[1]:SWEep:TRIGger?
Sets the trigger source for the AM swept–sine waveform.
BUS Enables LAN triggering using the *TRG command.
IMMediate Enables immediate triggering of the sweep event.
EXTernal Enables the triggering of a sweep event by an externally applied signal at the
TRIGGER IN connector.
Refer to “:AM[1]|2:INTernal[1]|2:FUNCtion:SHAPe" on page 277 for the waveform
selection.
Example
:AM1:INT1:SWE:TRIG EXT
The example sets an external trigger source for the swept–sine waveform on AM path 1.
*RST IMM
:AM[1]|2:SOURce
Supported Option UNT
[:SOURce]:AM[1]|2:SOURce INT[1]|INT2|EXT[1]|EXT2