Technical data

268 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
8 Basic Function Commands
:STOP
Supported All
[:SOURce]:POWer:STOP <val><unit>
[:SOURce]:POWer:STOP?
Sets the amplitude of the last point in a step or ramp sweep (Option 007).
During an amplitude sweep, N8211A/N8212As with Option 1E1 protect the step attenuator
by switching to attenuator hold (ON) mode. The attenuator is locked at its current setting
and the amplitude sweep range is limited to 40 dB.
Example
:POW:STOP -10DBM
The example sets the amplitude of the last point in the sweep to –10 dBm.
*RST Depends on model and option number.
Range See “[:LEVel][:IMMediate][:AMPLitude]" on page 269 for the available power
ranges.
[:LEVel][:IMMediate]:OFFSet
Supported All
[:SOURce]:POWer[:LEVel][:IMMediate]:OFFSet <val><unit>
[:SOURce]:POWer[:LEVel][:IMMediate]:OFFSet?
Sets the power offset value as a dB power offset to the actual RF output. This simulates a
power level at a test point beyond the RF OUTPUT connector without changing the actual
RF output power. The offset value only affects the displayed amplitude setting.
You can enter an amplitude offset anytime in either normal operation or amplitude reference
mode.
Example
:POW:OFFS 10DB
The example sets the amplitude offset to 10 dB.
*RST +0.00000000E+000
Range 200 dB to 200 dB