Technical data
262 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
8 Basic Function Commands
Sets the stop frequency for a power search over a user-defined range. The stop frequency
has no default value. The stop frequency value will be saved before powering off the
instrument.
Example
:POW:ALC:SEAR:SPAN:STOP 20GHZ
The example selects 20 GHz as the stop frequency for a power search.
:ALC:SEARch:SPAN:TYPE FULL|USER
Supported All
[:SOURce]:POWer:ALC:SEARch:SPAN:TYPE FULL|USER
[:SOURce]:POWer:ALC:SEARch:SPAN:TYPE?
Selects the frequency range for a power search. You can specify the range (USER) or you
can select the full range (FULL) of the N8211A/N8212A.
Example
:POW:ALC:SEAR:SPAN:TYPE USER
The example selects a user-defined frequency range for the power search.
:ALC:SEARch:SPAN[:STATe] ON|OFF|1|0
Supported All
[:SOURce]:POWer:ALC:SEARch:SPAN[:STATe] ON|OFF|1|0
[:SOURce]:POWer:ALC:SEARch:SPAN[:STATe]?
Enables (1) or disables (0) the span mode, allowing you to perform power searches over a
selected range of frequencies. The power search corrections are then stored and used
whenever the N8211A/N8212A is tuned within the selected range.
Example
:POW:ALC:SEAR:SPAN ON
The example enables the span mode.
:ALC:SOURce
Supported All
[:SOURce]:POWer:ALC:SOURce INTernal|DIODe|MMHead