Technical data

Basic Function Commands 8
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 257
:MARKer:MODe
Supported Option 007
[:SOURce]:MARKer:MODE FREQuency|DELTa
[:SOURce]:MARKer:MODE?
Sets the frequency mode for all markers.
FREQuency The frequency values for the markers are absolute.
DELTa The frequency values for the markers are relative to the designated reference
marker. The reference marker must be designated before this mode is selected. (See
“:MARKer:REFerence" on page 257 to select a reference marker.)
Example
:MARK:MODE DELT
The example sets the marker mode to delta.
*RST FREQuency
:MARKer:REFerence
Supported Option 007
[:SOURce]:MARKer:REFerence <marker>
[:SOURce]:MARKer:REFerence?
Designates the reference marker when using markers in delta mode. The variable <marker>
designates the marker number.
Example
:MARK:REF 6
The example sets marker 6 as the reference marker.
*RST 0
Range 0-9
:MARKer[0,1,2,3,4,5,6,7,8,9][:STATe]
Supported Option 007
[:SOURce]:MARKer[0,1,2,3,4,5,6,7,8,9][:STATe] ON|OFF|1|0