Technical data

Basic Function Commands 8
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 255
Marker Subsystem–Option 007 ([:SOURce])
:MARKer:AMPLitude[:STATe]
Supported Option 007
[:SOURce]:MARKer:AMPLitude[:STATe] ON|OFF|1|0
[:SOURce]:MARKer:AMPLitude[:STATe]?
Sets the amplitude marker state for the currently activated markers. When the state is
switched on, the RF output signal exhibits a spike with a magnitude relative to the power
level at each marker’s set frequency. (To set the magnitude of the spike, refer to
“:MARKer:AMPLitude:VALue" on page 255.) The width of the amplitude spike is a nominal
eight buckets, based on 1601 buckets per sweep.
Example
:MARK:AMPL ON
The example enables amplitude markers.
*RST 0
:MARKer:AMPLitude:VALue
Supported Option 007
[:SOURce]:MARKer:AMPLitude:VALue <num>[DB]
[:SOURce]:MARKer:AMPLitude:VALue?
Sets the relative power for the amplitude spikes at each marker’s set frequency when the
amplitude marker mode is activated. (To activate the amplitude markers, refer to
“:MARKer:AMPLitude[:STATe]" on page 255.)
Example
:MARK:AMPL:VAL 4DB
The example sets the relative marker power to 4 dB for all markers.
*RST 2 dB
Range 10 dB to +10 dB