Technical data
Basic Function Commands 8
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 253
Example
:SWE:POIN 2001
The example sets the number of step sweep points to 2001.
:
SWEep:TIME
Supported Option 007
[:SOURce]:SWEep:TIME <val><units>
[:SOURce]:SWEep:TIME?
Sets the sweep time for a ramp sweep in seconds. If this command is executed while the
signal generator is in automatic sweep time mode, the manual sweep time mode is
activated and the new sweep time value is applied. The sweep time cannot be set to a value
faster than what the automatic mode provides.
The sweep time is the duration of the sweep from the start frequency to the stop frequency.
It does not include the bandcross time that occurs during a sweep or the retrace time that
occurs between sweep repetitions.
Example
:SWE:TIME .250
The example sets the ramp sweep time to 250 mS.
*RST 1.00000000E−002
Range 10 mS–99 S
:SWEep:TIME:AUTO
Supported Option 007
[:SOURce]:SWEep:TIME:AUTO ON|OFF|0|1
[:SOURce]:SWEep:TIME:AUTO?
Sets the sweep time mode for a ramp sweep.
The sweep time is the duration of the sweep from the start frequency to the stop frequency.
It does not include the bandcross time that occurs during a sweep or the retrace time that
occurs between sweep repetitions.
*RST 2
Range 2–65535