Technical data
248 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
8 Basic Function Commands
:LIST:POWer:POINts
Supported All
[:SOURce]:LIST:POWer:POINts?
Queries the number of power points in the current list sweep file.
:LIST:RETRace
Supported All
[:SOURce]:LIST:RETRace ON|OFF|1|0
[:SOURce]:LIST:RETRace?
Upon completion of a single sweep operation, this command either resets the sweep to the
first sweep point, or leaves it at the last sweep point. The command is valid for the list, step,
or ramp (Option 007) single-sweep modes.
ON (1) The sweep resets to the first sweep point.
OFF (0) The sweep stays at the last sweep point.
Example
:LIST:RETR 1
The example sets the retrace on. The sweep will reset to the first point after completing a
sweep.
*RST 1
:LIST:TRIGger:SOURce
Supported All
[:SOURce]:LIST:TRIGger:SOURce BUS|IMMediate|EXTernal
[:SOURce]:LIST:TRIGger:SOURce?
Sets the trigger source for a list or step sweep event.
To set the sweep trigger, see “:TRIGger[:SEQuence]:SOURce" on page 222.
BUS This choice enables LAN triggering using the *TRG command.
IMMediate This choice enables immediate triggering of the sweep event.