Technical data
244 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
8 Basic Function Commands
Example
:LIST:DIR UP
The example selects an ascending sweep direction.
*RST UP
:LIST:DWELl
Supported All
[:SOURce]:LIST:DWELl <val>{,<val>}
[:SOURce]:LIST:DWELl?
Sets the dwell time for points in the current list sweep.
The variable <val> is expressed in units of seconds with a 0.001 resolution. If only one point
is specified, that value is used for all points in the list. Otherwise, there must be a dwell
point for each frequency and amplitude point in the list.
The dwell time <val> does not begin until the N8211A/N8212A frequency and/or
amplitude change has settled.
Dwell time is used when IMMediate is the trigger source. Refer to
“:LIST:TRIGger:SOURce" on page 248 for the trigger setting.
The dwell time is the amount of time the sweep pauses after setting the frequency and/or
power for the current point.
The setting enabled by this command is not affected by a power cycle, preset, or *RST
command.
Example
:LIST:DWEL .1,.2,.1,.2,.3
The example sets the dwell time for a list of five points.
Range 0.001–60
:LIST:DWELl:POINts
Supported All
[:SOURce]:LIST:DWELl:POINts?
Queries the N8211A/N8212A for the number of dwell points in the list sweep file.