Technical data
240 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
8 Basic Function Commands
[:SOURce]:FREQuency[:CW]:STEP[:INCRement] <val><unit>
[:SOURce]:FREQuency[:CW]:STEP[:INCRement]?
Sets the incremental step value for the frequency parameter. The value set with this
command is not affected by *RST or a power cycle.
Range .01 Hzā99 GHz
:PHASe:REFerence
Supported All
[:SOURce]:PHASe:REFerence
Sets the output phase reference to zero. Subsequent phase adjustments are set relative to
the new reference.
:PHASe[:ADJust]
Supported All
[:SOURce]:PHASe[:ADJust] <val><unit>
[:SOURce]:PHASe[:ADJust]?
Adjusts the phase of the modulating signal. The query returns values in radians.
Example
:PHAS 30DEG
The example sets the phase of the modulating signal to 30 degrees relative to the previous
phase setting.
*RST +0.00000000E+000
Range Radians: ā3.14 to 3.14 rad, Degrees: ā180 to 179 deg
:ROSCillator:BANDwidth:DEFaults
Supported Option UNR/UNX
[:SOURce]:ROSCillator:BANDwidth:DEFaults
Resets the bandwidth of the reference oscillator to the factory-defined default state. The
default value for the internal reference bandwidth is 125 Hz. The default value for the
external reference bandwidth is 25 Hz.