Technical data
Basic Function Commands 8
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 239
:FREQuency[:CW]
Supported All
[:SOURce]:FREQuency[:CW] <val><unit>|UP|DOWN
[:SOURce]:FREQuency[:CW]?
Sets the N8211A/N8212A output frequency for the CW frequency mode, or increments or
decrements the current RF frequency setting.
<val> A frequency value.
UP Increases the current frequency setting by the value set with the
“:FREQuency[:CW]:STEP[:INCRement]" on page 239.
DOWN Decreases the current frequency setting by the value set with the
“:FREQuency[:CW]:STEP[:INCRement]" on page 239.
To set the frequency mode to CW, refer to “:FREQuency:MODE" on page 234.
Example
:FREQ 12GHZ
The example sets N8211A/N8212A’s output frequency to 12 GHz.
:FREQuency[:CW]:STEP[:INCRement]
Supported All
*RST Option 520: +2.0000000000000E+10
Option 540
*
: +4.0000000000000E+10
* N8211A only
Range Option 520: 250kHZ–20GHZ
Option 540
*
: 250kHZ–40GHZ
*RST Option 520: +2.0000000000000E+10
Option 540
*
: +4.0000000000000E+10
* N8211A only
Range Option 520: 250kHZ–20GHZ
Option 540
*
: 250kHZ–40GHZ