Technical data
232 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
8 Basic Function Commands
Key Entry Channel Number
:FREQuency:CHANnels[:STATe]
Supported All
[:SOURce]:FREQuency:CHANnels[:STATe] ON|OFF|1|0
[:SOURce]:FREQuency:CHANnels[:STATe]?
Enables or disables the frequency channel and band selection. The N8211A/N8212A
frequency will be set to the channel frequency when the state is on. To set frequency
channel bands refer to “:FREQuency:CHANnels:BAND" on page 229.
Example
:FREQ:CHAN ON
The example turns on the frequency channel.
*RST 0
:FREQuency:FIXed
Supported All
[:SOURce]:FREQuency:FIXed <val><unit>|UP|DOWN
[:SOURce]:FREQuency:FIXed?
800MHZ Base/Mobile: 0–640
1500MHZ Base/Mobile: 0–960
TETRA 380/390 Mobile: 3600–4000
TETRA 390/4000 Base: 3600–4000
TETRA 410/420 Mobile: 800–1200
TETRA 420/430 Base: 800–1200
TETRA 460/470: 2400 through 2800 2400–2800
TETRA 870/876 Mobile: 600–640
TETRA 915/921 Base: 600–940
PHS Standard: 1–255
DECT Standard: 0–9