Technical data

Basic Function Commands 8
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 229
Frequency Subsystem ([:SOURce])
:FREQuency:CENTer
Supported Option 007
[:SOURce]:FREQuency:CENTer <val>[<unit>]|UP|DOWN
[:SOURce]:FREQuency:CENTer? [MAXimum|MINimum]
Sets the center frequency for a ramp sweep. The center frequency symmetrically divides
the selected frequency span and is coupled to the start and stop frequency settings. The
frequency range and reset values are dependent on the signal generator model and option
number.
The query returns the start and stop ramp frequencies if the optional MAXimum or
MINimum are used.
Example
:FREQ:CENT 15GHZ
The example sets the center frequency for a ramp sweep to 15 GHz.
:FREQuency:CHANnels:BAND
Supported All
[:SOURce]:FREQuency:CHANnels:BAND
NBASe|NMOBile|BPGSm|MPGSm|BEGSm|MEGSm|BRGSm|MRGSm|BDCS|MDCS
|BPCS|MPCS|B450|GM450|B480|M480|B850|M850|B8|M8|B15|M15|B390|B420
|B460|B915|M380|M410|M450|M870|PHS|DECT
[:SOURce]:FREQuency:CHANnels:BAND?
Sets the frequency of the N8211A/N8212A by specifying a frequency channel band. The
frequency channel state must be enabled for this command to work. See
“:FREQuency:CHANnels[:STATe]" on page 232.
*RST Option 520: +2.0000000000000E+10
Option 540
*
: +4.0000000000000E+10
* N8211A only
Range Option 520: 250kHZ–20GHZ
Option 540
*
: 250kHZ–40GHZ
* N8211A Only