Technical data

Basic Function Commands 8
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 227
:FLATness:POINts
Supported All
[:SOURce]:CORRection:FLATness:POINts?
Returns the number of points in the user-flatness correction file.
:FLATness:PRESet
Supported All
Presets the user-flatness correction to a factory-defined setting that consists of one
frequency point and one amplitude point with no corrections.
Once this command is executed, correction data is overwritten; If needed, save the current
correction data (see “:FLATness:STORe" on page 227).
[:SOURce]:CORRection:FLATness:PRESet
:FLATness:STORe
Supported All
[:SOURce]:CORRection:FLATness:STORe "<file_name>"
Stores the current user-flatness correction data to a file named by the "<file_name>"
variable. All user-flatness files are stored in the N8211A/N8212A’s USERFLAT directory.
The directory path does not need to be specified in the command.
For information on file name syntax, refer to “File Name Variables" on page 164.
Example
:CORR:FLAT:STOR "New_Flat_data"
The example stores the current user-flatness table entries in a file named
"New_Flat_data".
[:STATe]
Supported All
[:SOURce]:CORRection[:STATe] ON|OFF|1|0
[:SOURce]:CORRection[:STATe]?
Toggles the application of user-flatness corrections to the current N8211A/N8212A power
output.