Technical data
226 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
8 Basic Function Commands
Correction Subsystem ([:SOURce]:CORRection)
:FLATness:LOAD
Supported All
[:SOURce]:CORRection:FLATness:LOAD "<file_name>"
Loads a user-flatness correction file designated by the file name "<file_name>" variable.
The file will be loaded from the N8211A/N8212A’s USERFLAT directory. The directory path
does not need to be specified in the command.
For information on file name syntax, refer to “File Name Variables" on page 164.
Example
:CORR:FLAT:LOAD "Flatness_Data"
The example loads a user flatness file named Flatness_Data from the N8211A/N8212A’s
user flatness directory.
:FLATness:PAIR
Supported All
[:SOURce]:CORRection:FLATness:PAIR <freq>,<corr>
Adds or edits a frequency and amplitude correction pair. The maximum number of pairs or
points that can be entered is 1601.
The <corr> variable is the power correction in dB.
Example
:CORR:FLAT:PAIR 10MHZ,.1
The example enters a frequency of 10 MHz and a power of 0.1 dB into the user flatness
table.
*RST Option 520: +2.0000000000000E+10
Option 540
*
: +4.0000000000000E+10
* N8211A only
Range Option 520: 250kHZ–20GHZ
Option 540
*
: 250kHZ–40GHZ