Technical data
System Commands 7
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 223
BUS Enables LAN triggering using the *TRG command.
IMMediate Enables immediate triggering of the sweep event.
EXTernal Enables the triggering of a sweep event by an externally applied signal at the
TRIGGER IN connector.
The wait for the BUS or EXTernal can be bypassed by sending the
:TRIGger[:SEQuence][:IMMediate]
command.
Example
:TRIG:SOUR BUS
The example sets the sweep trigger source to BUS.
*RST IMM
:TRIGger[:SEQuence][:IMMediate]
Supported All
:TRIGger[:SEQuence][:IMMediate]
Starts a List or Step sweep without the selected trigger occurring.
In some atypical cases, the :TRIG command could be ignored if it immediately follows
an*OPC? command. If the :TRIG command is ignored, then use a 10 ms sleep function
before sending the command.