Technical data

System Commands 7
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 217
The example selects a maximum length PN9 sequence.
*RST NORM
:PRESet:TYPE
Supported All
:SYSTem:PRESet:TYPE NORMal|USER
:SYSTem:PRESet:TYPE?
Selects the preset state for either factory or user-defined conditions. Refer to
:PRESet[:USER]:SAVE below for saving the USER choice preset settings. The setting
enabled by this command is not affected by a signal generator power-on, preset, or *RST
command.
Example
:SYST:PRES:TYPE USER
The example selects a user defined conditions for the signal generator preset state.
:PRESet[:USER]:SAVE
Supported All
:SYSTem:PRESet[:USER]:SAVE
Saves your user-defined preset conditions to a state file.
Only one user-defined preset file can be saved. Subsequent saved user–defined preset files
will overwrite the previously saved file.
:SECurity:ERASeall
Supported All
:SYSTem:SECurity:ERASall
This command removes all user files and flatness correction files. In addition, all table
editor files are returned to their original factory values.
This command differs from the :DELete:ALL command, which does not reset table editors to
factory values.
:SECurity:LEVel
Supported All