Technical data

System Commands 7
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 215
:OEMHead:FREQuency:MULTiplier
Supported All
:SYSTem:OEMHead:FREQuency:MULTiplier <val>
:SYSTem:OEMHead:FREQuency:MULTiplier?
Selects a multiplier for an external millimeter-wave source module.
The multiplier factor allows the N8211A/N8212As frequency display to show the source
module’s frequency. The selection is valid only when the OEM source module is selected.
The N8211A/N8212A’s actual RF frequency is not changed by the multiplier. For example, if
the N8211A/N8212A’s RF frequency is 20 GHz and a 4.000 x multiplier is selected, the
N8211A/N8212A will display 80 GHz.
Refer to the “:FREQuency:OFFSet" on page 235 and “:FREQuency:REFerence" on page 236
command descriptions for more information.
Example
:SYST:OEMH:FREQ:MULT 4
The example selects a 4x multiplier so that the N8211A/N8212A display shows the
frequency at the output of the mm-wave source module.
*RST 4.00000000E+000
:PON:TYPE
Supported All
:SYSTem:PON:TYPE PRESet|LAST
:SYSTem:PON:TYPE?
Sets the defined conditions for the N8211A/N8212A at power on.
PRESet Sets the conditions to factory- or user-defined as determined by the choice for
the preset type. Refer to “:PRESet:TYPE" on page 217 for selecting the type of preset.
LAST Retains the settings at the time the N8211A/N8212A was last powered down.
The selection is not affected by a power-on, preset, or the *RST command.
When LAST is selected, no N8211A/N8212A interaction can occur for at least 3 seconds
prior to cycling the power for the current settings to be saved.
Example
:SYST:PON:TYPE PRES