Technical data
214 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
7 System Commands
:SYSTem:OEMHead:SELect?
Selects an external millimeter-wave source module. The ON, REAR, and FRONT parameters
select the OEM source module while the OFF and NONE parameters deselect the OEM
source module.
Example
:SYST:OEMH:SEL ON
The example turns on the OEM source module.
*RST Off
:OEMHead:FREQuency:BAND WR15|WR12|WR10|WR8|WR6|WR5|WR3
Supported All
:SYSTem:OEMHead:FREQuency:BAND WR3
:SYSTem:OEMHead:FREQuency:BAND?
Selects a pre-defined waveguide rectangular (WR) frequency band. The WR selection is
determined by the external millimeter-wave source module frequency range. Selection of a
WR frequency band sets the minimum and maximum frequency bands, for the external
mm-wave source module, to pre-defined values shown in the table below. These
pre-defined frequency bands are common to commercially available mixers and multipliers.
Example
:SYST:OEMH:FREQ:BAND WR12
The example selects the 60-90 GHz WR frequency band.
*RST WR15
Table 17
Waveguide Band Start Frequency Stop Frequency Multiplier
WR15 50–75GHz 12.5000000000 GHz 18.7500000000 GHz 4.000 x
WR12 60–90GHz 10.0000000000 GHz 15.0000000000 GHz 6.000 x
WR10 75–110GHz 12.5000000000 GHz 18.4000000000 GHz 6.000 x
WR8 90–140GHz 11.2200000000 GHz 17.5000000000 GHz 8.000 x
WR6 110–170GHz 9.1000000000 GHz 14.2000000000 GHz 12.000 x
WR5 140–220GHz 11.6000000000 GHz 18.4000000000 GHz 12.000 x
WR3 220–325GHz 12.2000000000 GHz 18.1000000000 GHz 18.000 x