Technical data

System Commands 7
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 213
:OEMHead:FREQuency:STARt
Supported All
:SYSTem:OEMHead:FREQuency:STARt <val><units>
:SYSTem:OEMHead:FREQuency:STARt?
Sets the start frequency or minimum band frequency for an external source module. The
pre-defined start or minimum band frequency for the selected WR (waveguide rectangular)
is overwritten with this command. For more information on pre-defined frequency bands,
refer to “:OEMHead:FREQuency:BAND WR15|WR12|WR10|WR8|WR6|WR5|WR3" on
page 214.
Example
:SYST:OEMH:FREQ:STAR 90GHZ
The example sets the start frequency for the OEM module to 90 GHz.
*RST 5.0000000000000E+10
:OEMHead:FREQuency:STOP
Supported All
:SYSTem:OEMHead:FREQuency:STOP <val><units>
:SYSTem:OEMHead:FREQuency:STOP?
Sets the stop frequency or maximum band frequency for an external source module. The
pre-defined stop or maximum band frequency for the selected WR (waveguide rectangular)
is overwritten with this command. For more information on pre-defined frequency bands,
refer to “:OEMHead:FREQuency:BAND WR15|WR12|WR10|WR8|WR6|WR5|WR3” on
page 89.
Example
:SYST:OEMH:FREQ:STOP 70GHZ
The example sets the stop frequency for the OEM module to 70 GHz.
*RST 7.0000000000000E+10
:OEMHead:SELect
Supported All
:SYSTem:OEMHead:SELect ON|OFF|NONE|REAR|FRONT