Technical data

System Commands 7
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 211
System Subsystem (:SYSTem)
:ALTernate
Supported Option 007
:SYSTe m:ALTernate <reg _ nu m>
:SYSTe m:ALTernate ? [ M AXim u m|MI N i mum]
Sets up the N8211A/N8212A to use a sweep state stored in a state register to alternate
with the current sweep. The alternate sweep state must be stored in state registers 1
through 9 in sequence 0. Alternate sweep must be selected and both sweeps must be ramp
sweeps.
Example
:SYST:ALT 3
The example alternates the current sweep with the sweep settings saved in state register
number three.
:ALTernate:STAte
Supported Option 007
:SYSTe m:ALTernate :STATe O N |OFF|1 |0
:SYSTem:STAte?
Enables or disables the alternate sweep state for the N8211A/N8212A. With alternate
state on, the N8211A/N8212A uses the current sweep setup and alternates with a sweep
saved in on of the state registers. Both sweeps must be ramp sweeps.
Example
:SYST:ALT:STAT OFF
The example disables the alternate sweep mode.
:CAPability
Supported All
:SYSTem:CAPability?
Returns the N8211A/N8212A’s capabilities and outputs the appropriate specifiers:
RFSOURCE WITH((AM|FM|PULM|PM|LFO)&(FSSWEEP|FLIST)&(PSSWEEP|PLIST)
&TRIGGER&REFERENCE))