Technical data
210 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
7 System Commands
Example
:STAT:QUES:PTR 8
This example enables bit 3 (decimal 8, Power Summary) in the Data Questionable Positive
Transition Filter register.
Range 0–32767
:QUEStionable[:EVENt]
Supported All
This is a destructive read. The data in the register is latched until it is queried. Once queried,
the data is cleared.
:STATus:QUEStionable[:EVENt]?
This query returns the decimal sum of the bits in the Standard Operation Event register.
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
Range 0–32767