Technical data

208 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
7 System Commands
Enables bits in the Data Questionable Power Event Enable register. Bits enabled and set in
this register will set bit 3 in the Data Questionable Condition register.
The variable <val> is the sum of the decimal values of the bits that you want to enable.
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
Example
:STAT:QUES:POW:ENAB 1
This example enables bit 0 (decimal 1, Reverse Power Protection Tripped) in the Data
Questionable Power Event Enable register.
Range 0–32767
:QUEStionable:POWer:NTRansition
Supported All
:STATus:QUEStionable:POWer:NTRansition <val>
:STATus:QUEStionable:POWer:NTRansition?
Enables bits in the Data Questionable Power Negative Transition Filter register. A negative
transition (1 to 0) of corresponding bits in the Data Questionable Power Condition register
will pass through and be read by the Data Questionable Power Event register.
The variable <val> is the sum of the decimal values of the bits that you want to enable.
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
Example
:STAT:QUES:POW:NTR 1
This example enables bit 0 (Reverse Power Protection Tripped) in the Data Questionable
Power Negative Transition Filter register.
Range 0–32767
:QUEStionable:POWer:PTRansition
Supported All
:STATus:QUEStionable:POWer:PTRansition <val>
:STATus:QUEStionable:POWer:PTRansition?