Technical data

System Commands 7
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 205
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
Range 0–32767
:QUEStionable:MODulation:ENABle
Supported All
:STATus:QUEStionable:MODulation:ENABle <val>
:STATus:QUEStionable:MODulation:ENABle?
Enables bits in the Data Questionable Modulation Event Enable register. Bits enabled and
set in this register will set bit 7 in the Data Questionable Condition register.
The variable <val> is the sum of the decimal values of the bits that you want to enable.
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
Example
:STAT:QUES:MOD:ENAB 20
This example enables bit 2 (decimal 4, Modulation 1 Overmod) and bit 4 (decimal 16,
Modulation Uncalibrated) in the Data Questionable Modulation Event Enable register.
Range 0–32767
:QUEStionable:MODulation:NTRansition
Supported All
:STATus:QUEStionable:MODulation:NTRansition <val>
:STATus:QUEStionable:MODulation:NTRansition?
Enables bits in the Modulation Questionable Negative Transition Filter register. A negative
transition (1 to 0) of corresponding bits in the Modulation Questionable Condition register
will pass through and be read by the Modulation Questionable Event register.
The variable <val> is the sum of the decimal values of the bits that you want to enable.
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
Example
:STAT:QUES:MOD:NTR 3