Technical data

204 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
7 System Commands
:STATus:QUEStionable:FREQuency:PTRansition <val>
:STATus:QUEStionable:FREQuency:PTRansition?
Enables bits in the Data Questionable Frequency Positive Transition Filter register. A
positive transition (0 to 1) of corresponding bits in the Data Questionable Frequency
Condition register will pass through and be read by the Data Questionable Frequency Event
register.
The variable <val> is the sum of the decimal values of the bits that you want to enable.
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
Example
:STAT:QUES:FREQ:PTR 8
This example enables bit 3 (decimal 8, Baseband 1 Unlocked) in the Data Questionable
Frequency Positive Transition Filter register.
Range 0–32767
:QUEStionable:FREQuency[:EVENt]
Supported All
This is a destructive read. The data in the register is latched until it is queried. Once queried,
the data is cleared.
:STATus:QUEStionable:FREQuency[:EVENt]?
This query returns the decimal sum of the bits in the Data Questionable Frequency Event
register.
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
Range 0–32767
:QUEStionable:MODulation:CONDition
Supported All
:STATus:QUEStionable:MODulation:CONDition?
Returns the decimal sum of the bits in the Data Questionable Modulation Condition register.
For example, if the modulation is uncalibrated (bit 4), a value of 16 is returned.
The data in this register is continuously updated and reflects current conditions.