Technical data

202 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
7 System Commands
Range 0–32767
:QUEStionable:ENABle
Supported All
:STATus:QUEStionable:ENABle <val>
:STATus:QUEStionable:ENABle?
Enables bits in the Data Questionable Event Enable register. Bits enabled and set in this
register will set the Data Questionable Summary bit (bit 3) in the Status Byte register. When
bit 3 in the Status Byte register is set, you can read the Data Questionable Event register to
determine the cause.
The variable <val> is the sum of the decimal values of the bits that you want to enable.
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
Example
:STAT:QUES:ENAB 8
This example enables bit 3 (decimal 8, the Power Summary bit), in the Data Questionable
Event Enable register.
Range 0–32767
:QUEStionable:FREQuency:CONDition
Supported All
:STATus:QUEStionable:FREQuency:CONDition?
Returns the decimal sum of the bits in the Data Questionable Frequency Condition register.
For example, if the 1 GHz internal reference clock is unlocked (bit 2), a value of 4 is returned.
The data in this register is continuously updated and reflects current conditions.
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
Range 0–32767
:QUEStionable:FREQuency:ENABle
Supported All
:
STATus:QUEStionable:FREQuency:ENABle <val>