Technical data

198 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
7 System Commands
:OPERation:NTRansition
Supported All
:STATus:OPERation:NTRansition <val>
:STATus:OPERation:NTRansition?
Enables bits in the Standard Operation Negative Transition Filter register. A negative
transition (1 to 0) of corresponding bits in the Standard Operation Condition register will
pass through and be read by the Standard Operation Event register.
The variable <val> is the sum of the decimal values of the bits that you want to enable.
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
Example
:STAT:OPER:NTR 3
This command enables bit 0 (decimal 1, I/Q Calibrating) and bit 1 (decimal 2, Settling) in the
Standard Operation Negative Transition Filter register.
Range 0–32767
:OPERation:PTRansition
Supported All
:STATus:OPERation:PTRansition <val>
:STATus:OPERation:PTRansition?
Enables bits in the Standard Operation Positive Transition Filter register. A positive
transition (0 to 1) of corresponding bits in the Standard Operation Condition register will
pass through and be read by the Standard Operation Event register.
The variable <val> is the sum of the decimal values of the bits that you want to enable.
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
Example
:STAT:OPER:PTR 3
This command enables bit 0 (decimal 1, I/Q Calibrating) and bit 1 (decimal 2, Settling) in the
Standard Operation Positive Transition Filter register.
Range 0–32767