Technical data

System Commands 7
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 181
*RST
Supported All
*RST
Resets most N8211A/N8212A functions to a factory-defined state.
Each command description in this reference shows the *RST value if the
N8211A/N8212A’s setting is affected.
*SAV
Supported All
*SAV <reg>,<seq>
Saves the state of the N8211A/N8212A to the specified memory register <reg> of the
specified sequence <seq>. Settings such as frequency, attenuation, power, and settings
that do not survive a power cycle or an instrument reset can be saved. Data formats, arb
setups, list sweep values, table entries, and so forth are not stored. Only a reference to the
data file name is saved.
Range registers: 0–99 Sequences: 0–9
*SRE
Supported All
*SRE <val>
Enables bits in the Service Request Enable register. Bits enabled and set in this register will
set bits in the Status Byte register.
The variable <val> is the decimal sum of the bits that are enabled. Bit 6 (value 64) is not
available in this register and therefore cannot be enabled by this command. Because bit 6 is
not available, entering values from 64 to 127 is equivalent to entering values from 0 to 63.
Refer to Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
The setting enabled by this command is not affected by preset or *RST. However, cycling
the N8211A/N8212A power will reset this register to zero.
Range 0–63, 128–191
*SRE?
Supported All