Technical data
System Commands 7
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 171
• Path settings (Internal I/Q Mux Path 1 or Path 2)
• I/Q calibration (the I/Q DC calibration will be invalidated if any other I/Q calibration is
execute)
• Temperature (±5 degrees)
The following instrument states will not invalidate the I/Q DC calibration:
• Power level changes
• I/Q Impairments
*RST N/A
:IQ:DEFault
Supported N8212A
:CALibration:IQ:DEFault
Restores the original factory calibration data for the internal I/Q modulator.
:IQ:FULL
Supported N8212A
:CALibration:IQ:FULL
Sets and performs a full-frequency range (regardless of the start and stop frequency
settings) I/Q calibration and stores the results in the N8212A’s memory.
:IQ:STARt
Supported N8212A
:CALibration:IQ:STARt <val><units>
:CALibration:IQ:STARt?
Sets the start frequency and automatically sets the calibration type to User for an I/Q
calibration.
The setting enabled by this command is not affected by a power-on, preset, or *RST
command.
Example
:CAL:IQ:STAR 1GHZ
The example sets the start frequency for an IQ calibration to 1 GHz.