Technical data

Programming the Status Register System 4
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 129
Data Questionable Calibration Event Enable Register
The Data Questionable Calibration Event Enable Register lets you choose which bits in the
Data Questionable Calibration Event Register set the summary bit (bit 8 of the Data
Questionable Condition register) to 1.
Command: STATus:QUEStionable:CALibration:ENABle <value>, where <value> is
the sum of the decimal values of the bits you want to enable.
Example: Enable bit 1 and bit 0 so that whenever either of those bits are set to 1, the Data
Questionable Calibration summary bit of the Data Questionable Condition Register is set
to 1. Send the command STAT:QUES:CAL:ENAB 520 (2 + 1).
Query: STATus:QUEStionable:CALibration:ENABle?
Response: Decimal value of the sum of the bits previously enabled with the
STATus:QUEStionable:CALibration:ENABle <value> command.