Technical data

Programming the Status Register System 4
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 121
Data Questionable Power Event Enable Register
The Data Questionable Power Event Enable Register lets you choose which bits in the Data
Questionable Power Event Register set the summary bit (bit 3 of the Data Questionable
Condition Register) to 1.
Data Questionable Frequency Status Group
Some bits in this status group do not apply to the N8211A/N8212A and returns zero when
queried. See Ta b le 9 on page 122 for more information.
The Data Questionable Frequency Status Group is used to determine the specific event that
set bit 5 in the Data Questionable Condition Register. This group consists of the Data
Questionable Frequency Condition Register, the Data Questionable Frequency Transition
Filters (negative and positive), the Data Questionable Frequency Event Register, and the
Data Questionable Frequency Event Enable Register.
Command: STATus:QUEStionable:POWer:ENABle <value> where <value> is the sum of
the decimal values of the bits you want to enable
Example: Enable bit 3 and bit 2 so that whenever either of those bits are set to 1, the Data
Questionable Power summary bit of the Data Questionable Condition Register is set to 1.
Send the command STAT:QUES:POW:ENAB 520 (8 + 4).
Query: STATus:QUEStionable:POWer:ENABle?
Response: Decimal value of the sum of the bits previously enabled with the
STATus:QUEStionable:POWer:ENABle <value> command.