Technical data

Programming the Status Register System 4
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 119
The Data Questionable Power Status Group is used to determine the specific event that set
bit 3 in the Data Questionable Condition Register. This group consists of the Data
Questionable Power Condition Register, the Data Questionable Power Transition Filters
(negative and positive), the Data Questionable Power Event Register, and the Data
Questionable Power Event Enable Register.
Data Questionable Power Condition Register
The Data Questionable Power Condition Register continuously monitors the hardware and
firmware status of the N8211A/N8212A. Condition registers are read only.